| Model Number |
LTF-S190-10 |
| Medical Device Problem Codes |
No Display/Image (1183); Peeled/Delaminated (1454); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunctions: degradation and an electrical component issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the endoeye flex deflectable videoscope exhibited peeling glue at the objective lens and a black image with a scope communication error.There was no patient involvement.
|
| |
|