4110: a lens work order search has been performed: no similar complaints within associated lots were found.Manufacture narrative: secondary surgical intervention to reposition, remove, and replace the lens are identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced low vault with rotation.The lens was later reponsitioned.Reportedly, "toric lens rotated out of position," and "repostioned at different asc." the lens was later exchanged for a longer length lens and the problem was resolved.Cause of the event was other.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|