| Catalog Number |
8065000485 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Improper Flow or Infusion (2954); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported that stopped pumping air into the eye and produced an unusual noise during fluid air exchange for vitrectomy surgery.The surgery was completed on the same day.There was no report of patient impact.This report pertains to cassette involved in this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.6.And h.11.The product was not received at the investigating site for this complaint report; visual inspection or functional testing could not be conducted.The customer did provide a photo of the console screen with several system message codes; however, without the product we are unable to confirm root cause.The investigation conducted a non-conformance review of the reported lot number.No relevant deviations were identified, all corresponding production release specifications defined in the device master record were met.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.No further investigative or manufacturing actions are warranted at this time due to the product not being returned to the manufacturing site for testing.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|