This report has been identified as b.Braun medical internal report number (b)(4).One (1) used, not contaminated, sample without packaging was provided for evaluation.The sample was leak tested and no leakage was detected.To replicate the reported defect of the air in line alarm on the pump, the sample was attached to the pump and then tested by running therapy while staggering the flow rate throughout the therapy.No alarms occurred during the testing of the returned sample.Thereafter, a measurement of the outer diameter of the pump segment tubing was taken and found to be 0.159 inches, which meets the specification of 0.1556-0.1633 inches.Based on the investigation findings, the sample met internal specification; therefore, the reported defect was not confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|