MAUDE Adverse Event Report: CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION Back to Search Results
Model Number 8110
Medical Device Problem Codes Break (1069); Contamination (1120); Deformation Due to Compressive Stress (2889); Calibration Problem (2890); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device had drive arm slips during force accuracy.There was no patient involvement.
 
Additional Manufacturer Narrative
A follow up report will be submitted with investigation results should the device be repaired or the device/logs be received for evaluation.Per 803.52(f)(11)(iii) the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
 
Additional Manufacturer Narrative
A device history record, complaint history review, and risk review on failure modes are performed on each device with reportable malfunction(s) along with other methods of investigation as coded in section h6 of this mdr report.This report lists imdrf annex a, c, d, and g codes that are reportable malfunctions observed on the device during servicing.Per 803.52(f)(11)(iii), the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
 
Event or Problem Description
It was reported that the device had drive arm slips during force accuracy.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYSTEM
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
brett wilko
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key24765020
Report Number2016493-2026-18636
Device Sequence Number15741547
Product Code FRN
UDI-Device Identifier10885403424267
UDI-Public(01)10885403424267
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K133532
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number8110
Device Catalogue Number8110 ALARIS SYRINGE MODULE
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Supplement Date Received by Manufacturer04/21/2026
Initial Report FDA Received Date04/02/2026
Supplement Report FDA Received Date04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/12/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2719-2020
Patient Sequence Number1
Patient SexUnknown
-
-