| Model Number |
72404252 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 720185-01.Serial number: n/a.Batch/lot number: 0182467011.Model/catalog description: reservoir flat iz 100 ml.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Fluid leak and inflation issue are acknowledged within the dfu and are not related to off-label use or failure to follow instructions.A risk review was completed; fluid loss and inflation issue are defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with an inflatable penile prosthesis (ipp) was unable to inflate the device due to fluid loss.A surgical procedure was performed to remove and replace the ipp.No patient complications were reported.
|
| |
|