| Medical Device Problem Code |
Increase in Pressure (1491)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an error during pre-use checkout resulting in potential over delivery of pressure.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.Moisture was removed the tubing to resolve the issue.
|
| |
|