| Model Number |
M00558650 |
| Medical Device Problem Codes |
Inflation Problem (1310); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pro wireguided dase dilatation balloon was attempted to be used in the esophagus during a gastroscopy procedure for esophageal stricture performed on (b)(6) 2026.During the procedure, after inflation attempt, the medium was leaking from the balloon, creating inflation issues.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fully recovered.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0504 captures the reportable event of balloon leak in the esophagus block h11: investigation results: the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.With all available information, boston scientific concludes the reported event of balloon leak was not confirmed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.
|
| |
|
Search Alerts/Recalls
|