This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: investigation summary the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that the device comes in used condition as expected in reusable devices.The upper jaw was found slightly bent at the tip, which causes the jaw to not close completely.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the expressew iii ac+ gun would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life.As per instructions for use (ifu); visually inspect the instrument and check for damage and wear.Make sure that: jaws and teeth are aligned properly.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
|
It was reported that during in-house engineering evaluation, it was determined that the expressew iii ac+ gun device was deformed/bent.It was initially reported that during an unspecified surgical procedure, it was discovered that the top jaw did not close/fit properly on bottom jaw.There were no delays to the surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|