| Model Number |
61355009 |
| Medical Device Problem Code |
Electrical Overstress (2924)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A corrective and preventative action capa: 01205 has been initiated to thoroughly investigate the issue and identify the root cause.
|
| |
|
Event or Problem Description
|
|
The blower was sent to agiliti for repair with the reported issue of power failure.Upon opening the device, an agiliti technician discovered internal burn damage that caused the power failure.A burn mark approximately 9 inches by 7 inches was observed on the percussion power supply board extending to the back case and side panel of the blower.
|
| |
|