| Catalog Number |
368685 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There were multiple medical device types reported to be involved.The information for the additional device type is as follows.D.2.B medical device type: jka.D.2.A common device name: tubes, vials, systems, serum separators, blood collection.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
Report 2 of 2: it was reported while using bd vacutainer® ultratouch¿ push button blood collection set with pah, blood from an unspecified number of devices leaked between the tube and the holder.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received any samples or photos for investigation.Similarly, functional tests were conducted on 30 retained samples from batch number 5282522, and all samples passed as there was no evidence of side-pierced sleeves, poor sleeve recovery, sleeve fall off, or sleeve leakage during use.Based on a review of the device history record for the incident lots, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the lot 5282522, for the indicated failure mode: sleeve function.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
Report 2 of 2: it was reported while using bd vacutainer® ultratouch¿ push button blood collection set with pah, blood from an unspecified number of devices leaked between the tube and the holder.No adverse impact was reported regarding the patient or user.
|
| |
|