| Catalog Number |
130760000 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent the revision surgery due to fracture of the glenosphere´s screw.Based on the patient's history and imaging, the material failure appears to have been caused by gross misplacement of the metaglene with insufficient fit between the glenosphere and metaglene.This resulted in years of abnormal notching, creating a mechanical lever on the glenosphere and ultimately leading to fracture of the coupling screw.Implantation date: (b)(6) 2024.Affected side: left side.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Added: d4 (expiration date), d10 (concomitant) h4.Corrected: d4 (primary udi).
|
| |
|
Search Alerts/Recalls
|