As reported through the lvis pas clinical study there are 3 ae's have been reported for subject (b)(6) at site usa-06.An ae has been reported for subject (b)(6) at site usa-06.Below ae was adjudicated on 10mar2026 as study device-related, study procedure-related study: lvis pas (retrospective study): ae #1 name: transient occlusion of right mca.Event onset date: 31-oct-2018.Event description: transient occlusion of rigth mca with restoration of normal flow after infusion therapy for thrombolysis.Date of index procedure: (b)(6) 2018 is the event a serious adverse event: no.Outcome: resolved without sequelae.Does the event meet criteria for an unanticipated adverse device effect: no.The patient outcome was based on discharge summary on (b)(6) 2018 patient worked with pt/ot and did well enough to be discharged home with supervision.Has headaches which will be managed conservatively.Below ae # 4 was adjudicated on (b)(6) 2026 as study device-related, study procedure-related.Study: lvis pas (retrospective study): ae #4 name: hypoperfusion.Event onset date: 14-mar-2019.Event description: hypoperfusion event related to right carotid occlusion.Presented as a minor stroke with minimal symptoms that pt felt was brought on by vigorous exercise.Physician diagnosed this is hypoperfusion.Date of index procedure: (b)(6) 2018.Is the event a serious adverse event: no.Outcome: resolved without sequelae.Does the event meet criteria for an unanticipated adverse device effect: no.Below ae#5 was adjudicated on 10mar2026 as study device-related, study procedure-related.Study: lvis pas (retrospective study).Ae # 5 name: neurological symptoms - potentially tia.Event onset date: 15-aug-2019.Event description: transient neurological symptoms suspicious for tia.Mri did not reveal any ischemia.Date of index procedure: (b)(6) 2018.Is the event a serious adverse event: no.Outcome: resolved without sequelae.Does the event meet criteria for an unanticipated adverse device effect: no the patient outcome for ae events 4&5: the patient came in for follow-up visit on 06dec2019 after restarting plavix to discuss his platelet response testing results.Patient normal response to plavix but does not respond to aspirin.As a result, he will stop his aspirin and will continue plavix at 75mg daily for life.Due to the low risk of aneurysm recurrence, patient was planned to have routine.Angiogram at 6, 12 and 24 months after the date of aneurysm coiling with mra follow up for years 3,4, and 5.Patient was exited from study on (b)(6) 2019.
|
|
The physical device was not available for evaluation to investigate if a condition existed that would have contributed to the reported event.Supplemental imaging was also not provided for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation further examine the cause of the reported event.This information may be updated if additional information is provided at a later date.Based on a review of the device¿s risk documentation, the reported event did not indicate there were the presence of any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination could not be performed as this information was not available at the time this investigation was performed.Complaint system review: a search of the complaint handling system could not be performed to determine if other similar complaints exist for this batch number, because the batch number was not provided for the product on this complaint file.Ifu review (additional information can be found in the ifu): potential adverse events the following potential risks and complications associated with general anesthesia, cerebral angiography, intracranial catheterization, intracranial stent placement or intra-saccular coil deployment have been identified below: ¿ allergic reaction, including but not limited to: contrast dye, nitinol metal, and any other medications used during the procedure; ¿ aphasia ¿ blindness; ¿ cardiac arrhythmia; ¿ coil prolapsed or migration into normal vessel adjacent to aneurysm ¿ complications of arterial puncture including pain, local bleeding, local infection and injury to the artery, vein or adjacent nerves; ¿ cranial neuropathy; ¿ death; ¿ device fracture, migration or misplacement; ¿ dissection or perforation of the parent artery; ¿ headache; ¿ hemorrhage (i.E., intracerebral hemorrhage (ich), subarchnoid hemorrhage (sah), or retroperitoneal (or in other locations)); ¿ hemiplegia; ¿ hydrocephalus; ¿ infection; ¿ injury to normal vessel or tissue; ¿ ischemia; ¿ mass effect; ¿ myocardial infarction; ¿ neurological deficits; ¿ occlusion of non-target side branches; ¿ pseudo aneurysm formation; ¿ reactions to anti-platelet/anti-coagulant agents; ¿ reactions due to radiation exposure; ¿ reactions to anesthesia and related procedures; ¿ reactions to contrast agents; ¿ renal failure; ¿ aneurysm rupture; ¿ stenosis of stented segment; ¿ seizure; ¿ stent thrombosis; ¿ stroke or tia (transient ischemic attack); ¿ thromboembolic event (t/e); ¿ vasospasm; ¿ visual impairment.Warnings should unusual resistance be felt at any time during access or removal, the introducer/guide catheter/microcatheter and lvis device should be removed as a single unit.Applying excessive force during delivery or retrieval of the lvis device can potentially result in loss or damage to the device and delivery components.It is imperative to use the lvis device with compatible microcatheters.If repeated friction is encountered during lvis device delivery, verify microcatheter is not kinked or in extremely tortuous anatomy.Confirm that the microcatheter does not ovalize.Confirm that there is adequate sterile flush solution.Do not reposition the lvis device in the parent vessel without fully retrieving the device.The lvis device must be retrieved into the microcatheter and re-deployed at the desired target location or removed completely from the patient.Precautions exercise caution when crossing the deployed/detached lvis device with adjunctive devices such as guidewires, catheters, microcatheters or balloon catheters to avoid disrupting the device geometry and device placement.Directions for use 15.Advance the delivery wire to transfer the lvis device from within the introducer into the microcatheter.Warning: do not torque the delivery wire while advancing or retracting the lvis device.A torque device should not be used.16.Continue advancing the delivery wire into the microcatheter until the proximal tip of the delivery wire enters the introducer.Loosen the rhv locking ring, remove the introducer, and set it aside.Warning: do not apply undue force.If resistance is encountered at any point during lvis device delivery or manipulation, withdraw the unit and select a new lvis device.18.Position the lvis device for deployment by aligning the lvis implant distal radiopaque end markers approximately 7 mm past the aneurysm neck.Note: a proper push/pull technique, encompassing sufficient delivery wire push force, in addition to an opposing microcatheter withdrawal force, will facilitate properly deploying the lvis device to achieve full expansion and good vessel apposition.Note: slowly advancing the lvis device while adjusting the microcatheter position will ensure accurate deployment.Maintain simultaneous control of the lvis device and microcatheter in order to position and expand the device at the proper location.Caution: using a rapid microcatheter withdrawal technique to deploy the lvis device is not recommended and may result in device elongation.19.If lvis device positioning is not satisfactory, the lvis device may be recaptured and repositioned if it is not fully deployed.The lvis device may be recaptured until the point where the proximal end of the lvis device markers is aligned 3 mm proximally with the microcatheter distal marker band (approximately 80% deployed).Caution: if resistance is felt while recapturing the lvis device, do not continue to recapture the device.Withdraw the microcatheter slightly to unsheath the lvis device (without exceeding the recapture limit), and then attempt to recapture the lvis device.Caution: the lvis device must not be re-deployed more than three times.Note: the lvis device delivery wire should not be utilized as a guidewire after stent deployment.Do not torque the lvis device.A torque device should not be used.20.If lvis device positioning is satisfactory, carefully retract the microcatheter and advance the delivery wire together, to allow the lvis device to deploy across the neck of the aneurysm.Ensure the device proximal radiopaque end markers are approximately 7 mm proximal to the aneurysm neck to ensure an adequate landing zone.The lvis device will expand and total length may foreshorten up to 60% from its undeployed length as it exits the microcatheter.Ensure microcatheter is retracted and clear from the proximal flared ends.Note: visualize and refer to the implant radiopaque end markers to maintain adequate implant length, approximately 7 mm on each side of the aneurysm neck or target location to ensure appropriate neck coverage.Warning: do not detach the lvis device if it is not properly positioned in the parent vessel.Observe the delivery wire distal tip to assure it remains within the desired location of the parent vessel.21.Prior to removing the delivery wire and if necessary, carefully position the microcatheter distal to the lvis device to maintain access through the lvis device.Remove and discard the delivery wire.Warning: the lvis device delivery wire should not be utilized as a guidewire.Do not torque the lvis device.A torque device should not be used.
|