| Model Number |
3672 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's ipg became inoperable following an unrelated procedure.The patient did not set the ipg to surgery mode.Surgical intervention may be pending to address this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: event date is estimated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is included in the neuromodulation implantable pulse generator (ipg) proclaim¿ scs family, proclaim¿ drg therapy and infinity¿ dbs systems surgery mode advisory notice issued by abbott on 02 april 2026.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Additional information received reports that the patient underwent surgical intervention in which the ipg was explanted and replaced.A retrospective review indicates this event is connected to the neuromodulation implantable pulse generator (ipg) proclaim¿ scs family, proclaim¿ drg therapy and infinity¿ dbs systems surgery mode advisory notice issued by abbott on 02 april 2026.
|
| |
|