• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) IPG, 7.5AH, SCS; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) IPG, 7.5AH, SCS; SCS IPG Back to Search Results
Model Number 3672
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 03/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's ipg became inoperable following an unrelated procedure.The patient did not set the ipg to surgery mode.Surgical intervention may be pending to address this issue.
 
Additional Manufacturer Narrative
B3: event date is estimated.
 
Additional Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) proclaim¿ scs family, proclaim¿ drg therapy and infinity¿ dbs systems surgery mode advisory notice issued by abbott on 02 april 2026.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Additional information received reports that the patient underwent surgical intervention in which the ipg was explanted and replaced.A retrospective review indicates this event is connected to the neuromodulation implantable pulse generator (ipg) proclaim¿ scs family, proclaim¿ drg therapy and infinity¿ dbs systems surgery mode advisory notice issued by abbott on 02 april 2026.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPG, 7.5AH, SCS
Common Device Name
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
9723098000
MDR Report Key24772393
Report Number3006705815-2026-02067
Device Sequence Number15934091
Product Code LGW
UDI-Device Identifier05415067046406
UDI-Public(01)05415067046406(10)A000147743(17)250917
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other,Notification,Recall
Type of Report Initial,Followup
Report Date (Section B) 04/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2025
Device Model Number3672
Device Catalogue Number3672
Device Lot NumberA000147743
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2026
Supplement Date Received by Manufacturer04/15/2026
Initial Report FDA Received Date04/02/2026
Supplement Report FDA Received Date04/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1627487-04/02/26-001-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS ANCHOR (X2).; SCS LEAD (X2).
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight98 KG
-
-