|
W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT261412 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Obstruction/Occlusion (2422)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following was reported to gore: on (b)(6) 2026, the patient underwent endovascular treatment of an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprosthesis (the trunk ipsilateral leg endoprosthesis) and gore® excluder® aaa endoprosthesis devices.The patient tolerated the procedure.On (b)(6) 2026, a follow-up cta revealed the left renal artery obstruction due to a suspected proximal migration of the trunk ipsilateral leg endoprosthesis.On (b)(6) 2026, a reintervention was performed.A bare stent was implanted in the left renal artery.The patient tolerated the procedure.It was reported that, during the initial procedure, the trunk ipsilateral leg endoprosthesis had been implanted below the left renal artery and that there had been no issue with blood flow in the left renal artery.
|
| |
|
Search Alerts/Recalls
|
|
|