| Catalog Number |
0730615 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, the patient was prepped a catheter placement procedure using the powerline cv catheter prior to the procedure, it was reported that two triple lumen powerlines were opened and found to have damaged tunnelers.The end that is inserted into the catheter was bent, and when attempts were made to straighten them, the tunnelers snapped.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is inconclusive for the reported tunneler fracture, material separation and deformation due to compressive stress issues as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (medical device lot, unique identifier (udi) #), g3, h6 (method) section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, the patient was prepped a catheter placement procedure using the powerline cv catheter prior to the procedure, it was reported that two triple lumen powerlines were opened and found to have damaged tunnelers.The end that is inserted into the catheter was bent, and when attempts were made to straighten them, the tunnelers snapped.There was no patient contact.
|
| |
|