H11: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
|
It was reported and later learned that a patient with a 23mm 11500a aortic valve underwent a valve-in-valve procedure after an implant duration of 6 years due to stenosis.The patient presented with symptoms of heart failure, shortness of breath, and fatigue.Tavr was completed with a 26mm 9755rsl transcatheter valve and patient was in recovery post-operative.
|
|
H11: additional manufacturer narrative: updated sections: b4, g3, g6, h2, h6 (investigation findings, and investigation conclusions).The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: per (b)(4), stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hld and atherosclerosis.
|