H3 : product analysis :evaluation determined that detached bearings in the tube.The likely cause of the failure could be due to the tube loose from the housing therefore, a temperature test cannot be safely performed, it was noted also that the laser markings are faded and the output drive shaft's teeth are worn and the housing nut is deformed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|