| Catalog Number |
68-550 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported ¿saline rupture¿.This record is for the right side.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Clarification to d6a - partial implant date provided as 2003.Date in d6a was adjusted/removed as device was not manufactured until 26/apr/2005.Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification (if applicable).Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ deflation: observed two openings through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (creases and wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, h6.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued e1 -- alternate phone number:(b)(6).Continued e1 -- alternate email address:(b)(6).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: "saline rupture" (deflation).
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿saline rupture¿.This record is for the right side.Device remains implanted.
|
| |
|