| Model Number |
BF-XT190 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely that the following led to the malfunction: stress problem; cause traced to component failure.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
The bronchovideoscope was returned to the service center with a report of leak. this does not indicate an mdr reportable event. however, mdr reportable failure was found during testing at the service center.During inspection of the device, objective lens has chipped glue was found.There was no patient involvement.
|
| |
|