Model Number 0915612-06
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient aspirates and her pulmonary condition deteriorate.Given the overall situation, intubation should be avoided if possible; therefore, the decision is made to perform bronchoscopy under sedation via the nasal route.The endoscope becomes blocked several times due to very thick secretions and must be flushed repeatedly outside the patient.After approximately the third time, the cold light illumination at the front of the bronchoscope fails.The image from the camera is still visible on the monitor under room lighting, but as soon as the scope is inserted into the nostril, visibility is lost because there is no light.The bronchoscopy must be discontinued.The patient was reintubated following further pulmonary deterioration; however, this would likely have been necessary even if the bronchoscopy had been successful.Due to discontinuation of bronchoscopy and reintubation, a report is required.
 
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not returned at the time of mdr submission.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIVE S 5.3X65, STERILE, FOR SINGLE USE
Common Device Name
FIVE S 5.3X65, STERILE, FOR SINGLE USE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
MDR Report Key24779697
Report Number2020550-2026-00736
Device Sequence Number18625763
Product Code EOQ
UDI-Device Identifier04048551446690
UDI-Public04048551446690
Combination Product (Y/N)N
Initial Reporter CountrySZ
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/02/2026
Report Date (Section F) 04/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0915612-06
Device Catalogue Number0915612-06
Device Lot Number500101553
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date03/25/2026
Event Location Hospital
Date Report to Manufacturer04/02/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/03/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-