| Model Number |
FSESLIQ |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional fda product codes include: dsb- plethysmograph, impedance.Dxn- system, measurement, blood-pressure, non-invasive.Fll- continuous measurement thermometer.Mud- oximeter, tissue saturation.Qaq- adjunctive predictive cardiovascular indicator.Qem- cerebral oximeter.Qms- adjunctive open loop fluid therapy recommender.Qnl- medium-term adjunctive predictive cardiovascular indicator.The device evaluation is anticipated.A supplemental report will be forthcoming when the investigation is completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the foresight sensor the total hemoglobin was off by more than 1 g.There was no allegation of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Per review of the images provided by customer, it's confirmed image 1 shows a foresight thb of 9.2g/dl and image 2 shows arterial blood lab results with a hb of 6.8g/dl.The device was not returned for evaluation.Without the return of the unit, it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.A device history record review was unable to be completed as the lot number was not provided.
|
| |
|
Search Alerts/Recalls
|