• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 367336
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd received a photo for investigation.The indicated failure mode of leakage was observed in the attached photo.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: leakage.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Event or Problem Description
It was reported during colleciton using bd vacutainer® push button blood collection set, sample leakage from the barrel of one device was observed.No adverse impact was reported regarding the patient or user.
 
Event or Problem Description
It was reported during colleciton using bd vacutainer® push button blood collection set, sample leakage from the barrel of one device was observed.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
D2a.There were multiple medical device types reported to be involved.The information for the additional device type is as follows: jka.D2b.Common device name: tubes, vials, systems, serum separators, blood collection.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BECTON DICKINSON & CO (FRANKLIN LAKES)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
2018475186
MDR Report Key24781049
Report Number9617032-2026-00411
Device Sequence Number6132876
Product Code FPA
UDI-Device Identifier30382903673361
UDI-Public(01)30382903673361(11)250723(17)270731(10)5204764
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K220212
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number367336
Device Lot Number5204764
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2026
Supplement Date Received by Manufacturer04/03/2026
Initial Report FDA Received Date04/03/2026
Supplement Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-