• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUTRONIC CORPORATION XERF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUTRONIC CORPORATION XERF Back to Search Results
Catalog Number 1101300310
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 03/04/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On 3/18/2026 a trained cynosure lutronic field service engineer went to the treatment clinic for a device evaluation.During this time the device was subjected to full functionality testing, verified device set up and confirmed power measurement.The device was found to be working within published specifications.Visual inspection of the effector tip used during treatment observed a small black dot on the tips surface, which was reported to not be present prior to treatment.An image was taken of this damage, and its location suggests carbonization occurred at the ntc line tip lot and manufacturing date was checked and it was confirmed to be produced before the implementation of the resistance inspection that was completed in september 2025.Once the tip has been returned and evaluated, root cause can be better verified.A member of the cynosure lutronic clinical team contacted the treatment clinic for additional information.It was determined that the treatment parameters used were within the recommended clinical guidelines of the quickstart guide (qsg).It is undetermined what caused the reported burning, it is possible that the effector tip may not have had full contact with the treatment area when the burn occurred, though this cannot be confirmed.Burning is an expected side effect from treatment.The event is reportable per fda medical device reporting title 21 cfr part 803 since a malfunction occurred and would likely cause or contribute to a serious injury if the malfunction were to recur.We are reporting this as an initial mdr and will submit a final mdr once investigation has been completed.
 
Event or Problem Description
It was reported that a patient experienced a burn on the face during treatment performed on (b)(6) 2026 with a xerf device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XERF
Common Device Name
XERF
Manufacturer (Section D)
LUTRONIC CORPORATION
lutronic center
219 sowon-ro
deogyang-gu, goyang-si 10534
KS  10534
Manufacturer (Section G)
LUTRONIC CORPORATION
lutronic center
219, sowon-ro
deogyang-gu, goyang-si 10534
KS   10534
Manufacturer Contact
nirav desai
5 carlisle road
westford, MA 01886
MDR Report Key24781378
Report Number3021199089-2026-00019
Device Sequence Number9935456
Product Code GEI
UDI-Device Identifier08809447657981
UDI-Public(01)08809447657981(11)250429
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K050254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1101300310
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2026
Initial Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age31 YR
Patient SexFemale
-
-