MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-1TQ170
Medical Device Problem Codes Corroded (1131); Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was observed that during the device evaluation, the bronchovideoscope forceps channel port had corrosion.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection where a residue foreign material was confirmed.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24781676
Report Number9610595-2026-27308
Device Sequence Number15961715
Product Code EOQ
UDI-Device Identifier04953170342943
UDI-Public04953170342943
Combination Product (Y/N)N
PMA/510(K) Number
K121959
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-1TQ170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2026
Initial Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/09/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-