| Catalog Number |
21-7346-24 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient reached out yesterday regarding ongoing issues with air appearing in the bags after priming and preparation without any visible air however, after the bags sat for some time, visible air was observed.The patient believed the issue was related to the filter in the set.For hazardous doses, phaseal optima cstd was used.The tubing was primed via gravity.The recent increase in these occurrences was concerning, as no changes in practice had been made on their end and their technique had remained the same.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: codes updated.The suspected device was not received for evaluation.The device history file was reviewed.There were no nonconformities were identified that may have contributed to the reported complaint.The complaint cannot be replicated or confirmed, and a root cause was unable to be determined.A good faith effort was conducted to retrieve the device from the customer.
|
| |
|