It was reported that the bd maxzero minibore quad-fuse extension set was cracked.The following information was provided by the initial reporter: various bd extension sets ¿cracking¿, despite past education efforts.Trifuse and quadfuse microbore extension sets.There have been several products that have broken.Many days involved.Unsure what the lot numbers are.Packages get thrown out.This has been a problem since we received the product so it would span over several lot numbers.We are concerned that there is an increased risk of bloodstream infection.Additionally, when they crack it can be a choking hazard for pediatric patients who are touching their central line or piv.
|
|
Description of event
it was reported by customer that bd extension sets ¿cracking¿.Investigation results
the customer reported sets are cracking, and returned a bolus of samples under (b)(4), which were spread out among several investigations.There were 35 samples, 22 trifuse and 13 quadfuse.This complaint record will cover 6 of the quadfuse sets.Upon initial visual examination, the sets confirmed the report of component damage, with the male luer spin collar being cracked.In some cases, the collar was completely broken off the set.The samples were sent for further investigation to the supplier of the component.Device / batch history record (dhr) review could not be performed, as the batch number was not provided, preventing traceability to the applicable manufacturing records.The supplier of the component confirmed the failure, but was unable to define the root cause or trace it to their process.Due to the ongoing issue with the mz9283 quadfuse and mz9281 trifuse sets, the investigation has been escalated to a cross-functional team, and a funded project has been opened.The project will further investigate the issue.
|