Catalog Number MZ9283
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd maxzero minibore quad-fuse extension set was cracked.The following information was provided by the initial reporter: various bd extension sets ¿cracking¿, despite past education efforts.Trifuse and quadfuse microbore extension sets.There have been several products that have broken.Many days involved.Unsure what the lot numbers are.Packages get thrown out.This has been a problem since we received the product so it would span over several lot numbers.We are concerned that there is an increased risk of bloodstream infection.Additionally, when they crack it can be a choking hazard for pediatric patients who are touching their central line or piv.
 
Event or Problem Description
Samples investigation.
 
Additional Manufacturer Narrative
Description of event it was reported by customer that bd extension sets ¿cracking¿.Investigation results the customer reported sets are cracking, and returned a bolus of samples under pr 14414051, which were spread out among several investigations.There were 35 samples, 22 trifuse and 13 quadfuse.This complaint record will cover 6 of the quadfuse sets.Upon initial visual examination, the sets confirmed the report of component damage, with the male luer spin collar being cracked.In some cases, the collar was completely broken off the set.The samples were sent for further investigation to the supplier of the component.Device / batch history record (dhr) review could not be performed, as the batch number was not provided, preventing traceability to the applicable manufacturing records.The supplier of the component confirmed the failure, but was unable to define the root cause or trace it to their process.Due to the ongoing issue with the mz9283 quadfuse and mz9281 trifuse sets, the investigation has been escalated to a cross-functional team, and a funded project has been opened.The project will further investigate the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO MINIBORE QUAD-FUSE EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A DE C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX  22250
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A DE C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX   22250
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key24783712
Report Number9616066-2026-00891
Device Sequence Number9988026
Product Code FPA
UDI-Device Identifier10885403230660
UDI-Public(01)10885403230660
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMZ9283
Device Lot Number25039189
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2026
Initial Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1