MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO PGLA HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number PV-12-40-3D
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that a concerto coil could not be advanced.No patient symptoms or complications were associated with this event.It was unknown what condition was being treated or what the patient's blood flow and vessel tortuosity were.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).Additional information was received stating that the procedure was completed by using a new product.The cause of the resistance was not determined.
 
Additional Manufacturer Narrative
H3 product analysis: as found condition (condition of returned device): a concerto coil was returned for analysis within a shipping box; within sealed biohazard pouch and within an opened concerto implant coil inner pouch.The unknown micro catheter used in the event was not returned.The implant coil was returned without introducer sheath.Visual inspection/damage location details: the concerto coil pushwire was found to be broken at actuator interface and kinked at 159.4cm from proximal end.The implant coil was found to be already detached and not returned.All other subassemblies appeared to be normal, and no other anomalies were observed.Testing/analysis (including sem reports): the concerto coil could not be used for resistance testing with an in-house micro catheter due to its damaged condition.Conclusion: based on the device analysis and the reported information, the customers report of ¿coil resistance/stuck in cath¿ was unable to be confirmed.Possible causes for coil resistance in catheter are, but not limited to, use of an incompatible device for delivery, severely tortuous patient anatomy, failure to hydrate the concerto coil prior to use, and lack of continuous flush during delivery.The model or lot number for the unknown catheter were not provided; therefore, compatibility could not be assessed.The root cause could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCERTO PGLA HLX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key24783997
Report Number2029214-2026-00604
Device Sequence Number18623189
Product Code KRD
UDI-Device Identifier00847536045316
UDI-Public(01)00847536045316(17)261003(10)B647241
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K090046
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/03/2026
Device Model NumberPV-12-40-3D
Device Catalogue NumberPV-12-40-3D
Device Lot NumberB647241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Initial Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-