|
It was reported that the patient did not charge their ipg as recommended and the ipg became unable to communicate with external devices.Additionally, the patient experienced ineffective therapy.As a result, surgical intervention was undertaken wherein the entire system was replaced on (b)(6) 2026 to address the issue.It is unknown which adapter is responsible for this issue.
|
|
Date of event is estimated.Additional components potentially involved in the event include: common device name: 8-channel adapter, mdt, 10cm, model: 2311, udi: (b)(4), serial: (b)(6), batch: 6179467.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|