SMITH & NEPHEW, INC. INTERTAN 10S 10MM X 20CM 125D; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 71675385
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/30/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: internal complaint reference: case-(b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, after an internal fixation surgery had been performed on an unknown date, the patient experienced a distal break on one (1) intertan 10s 10mm x 20cm 125d.This adverse event was treated by a revision surgery on (b)(6) 2026, in which the nail was revised.Current health status of patient remain unknown.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.The visual inspection finds it fractured into two pieces.There are also scratches and scuffs, consistent with the explantation activities.A lab analysis revealed that the intramedullary (im) nail fractured by the initiation and subsequent propagation of fatigue cracking.The fatigue cracking eventually propagated to an extent that the remaining cross-sectional area of the nail could not bear the imposed patient loading, which led to an overload fracture.Fatigue cracking was likely caused by the nail bearing cyclic stresses more than the material endurance limit for an extended period of time.These excessive cyclic stresses may be caused by any number of conditions, including but not limited to excessive patient activity levels prior to full bone union, applications of loads in excess of the material¿s strength, and/or poor bone quality.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed a similar event for the listed part number over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for intramedullary nail system revealed in possible adverse effects that cracking or fracture of the implant may occur.Preoperative planning section states that proper type and size of implant must be selected to ensure effective treatment of patients considering factors such as patient¿s size, strength, skeletal characteristics, skeletal health, and general health.Overweight or musculoskeletal deficient or unhealthy patients may create greater loads on implants that may lead to breakage or other failure of the implants.Additionally, postoperative care section warns that early weight bearing substantially increases implant loading and increases the risk of breaking the device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According to inspection drawing, the final inspection includes the verification of part configuration per print.Also, according to material specification, the quality and manufacture of standard grade titanium-6 aluminum-4 vanadium alloy shall be controlled.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include size selected, postoperative care, patient anatomy, abnormal loading of limb, excessive forces and/or injury.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10S 10MM X 20CM 125D
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24786993
Report Number1020279-2026-00472
Device Sequence Number13258954
Product Code JDS
UDI-Device Identifier00885556343647
UDI-Public885556343647
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K241804
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71675385
Device Lot Number25DM08689
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Supplement Date Received by Manufacturer05/27/2026
Initial Report FDA Received Date04/05/2026
Supplement Report FDA Received Date06/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-