OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 6,4/7,8 FR. X 330 MM, 7°, STRAIGHT OCULAR ANGLE, 4,2 FR. CHANNEL
|
Back to Search Results |
|
| Model Number |
WA29049A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the most probable cause of the malfunction was traced to component failure.The legal manufacture reviewed the customer provided the cleaning disinfection and sterilization (cds) processes where no obvious deviations from instructions for use (ifu) were identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.Three attempts were performed to obtain additional information, but no response was received from the customer.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the ureteroscope exhibited chipped objective lens.There was no patient involvement.
|
| |
|