| Model Number |
WB991036 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The generator was returned to the service center with a report of the contact plate was not working.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, the return electrode cable loose and an error e232 was found.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for inspection.Based on the results of the investigation and the information provided it is likely due to polymer optical fiber (pof) being loose.The most probable cause is component failure.The root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|