OLYMPUS WINTER & IBE GMBH BERLIN FACILITY OLYMPUS ESG-100, 220...240 V~,; WB991036 Back to Search Results
Model Number WB991036
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
The generator was returned to the service center with a report of the contact plate was not working.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, the return electrode cable loose and an error e232 was found.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned for inspection.Based on the results of the investigation and the information provided it is likely due to polymer optical fiber (pof) being loose.The most probable cause is component failure.The root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OLYMPUS ESG-100, 220...240 V~,
Common Device Name
WB991036
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH BERLIN FACILITY
rheinstrasse 8
teltow, brandenburg 14513
GM  14513
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH BERLIN FACILITY
rheinstrasse 8
teltow, brandenburg 14513
GM   14513
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24789375
Report Number3003724334-2026-00203
Device Sequence Number8806192
Product Code GEI
UDI-Device Identifier04042768910364
UDI-Public04042768910364
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K073207
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberWB991036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/12/2026
Initial Report FDA Received Date04/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-