Catalog Number 510026522
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 03/17/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had an anatomic shoulder arthroplasty done on (b)(6) 2025.Patient stated they were doing well for 4-5 months until they suddenly couldn¿t lift their arm due to a subscap tear.Today, the patient was converted from an anatomic shoulder to a reverse shoulder.The anatomic humeral head was removed along with the offset taper adapter, the hybrid glenoid was removed and replaced with a reverse baseplate.Stemless component was removed and replaced with a standard stem.Doi: (b)(6) 2025.Dor: (b)(6) 2026.Affected side: unknown.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d1, d10 (concomitant).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYBRID UNITI GLENOID M 26.5
Common Device Name
SHOULDER IMPLANT - SHOULDER GLENOID
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
RESTORATION MEDICAL POLYMERS
4474 e. park 30 dr
columbia city IN
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24790327
Report Number1818910-2026-05981
Device Sequence Number8764121
Product Code HSD
UDI-Device Identifier10603295549154
UDI-Public(01)10603295549154(17)291031(10)MI177607
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K212933
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number510026522
Device Lot NumberMI177607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/18/2026
Supplement Date Received by Manufacturer04/08/2026
Initial Report FDA Received Date04/06/2026
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ANATOMIC OFFSET TAPER ADAPTER; HUMERAL HEAD 42X14.5; STEMLESS M 36
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-