| Catalog Number |
259807570 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11: additional manufacturer narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure it was observed that the tip of the retaining stem inserter device broke off.It was reported that the procedure was continued with another like device.There was no delay reported.No additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: tip broke off.The product returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis revealed that the distal threaded tip of the rod weldment was broken; fragment was not returned for evaluation.No other anomalies were observed.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces, such as, impaction forces while the device is not fully o properly assemble, heavy usage or excessive force in a prying motion.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection and functional testing were not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the retaining stem inserter would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
|
| |
|