| Catalog Number |
5532-G-611 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
|
| Date of Event |
03/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The following information was provided: patient was brought to or with slight instability and had clunking in the knee.At revision arthroplasty, it was discovered the post on a ps poly implant had broken off at the base.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event:
an event regarding crack/fracture involving a triathlon insert was reported.The event was confirmed via evaluation of the provided photographs.Method & results:
-product evaluation and results: visual inspection: the reported device was not returned however photographs were provided for review.The photographs show a recently explanted insert with evidence of the post having fractured off the device.Material analysis, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions:
it was reported that 'patient was brought to or with slight instability and had clunking in the knee.At revision arthroplasty, it was discovered the post on a ps poly implant had broken off at the base.'
the reported device was not returned however photographs were provided for review.The photographs show a recently explanted insert with evidence of the post having fractured off the device
no further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|