Break (1069); Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code
No Clinical Signs, Symptoms or Conditions (4582)
Date of Event
03/11/2026
Type of Reportable Event
Malfunction
Event or Problem Description
It was reported that during a remote follow up, noise resulting in oversensing and inappropriate automatic mode switch (ams) was noted on the right atrial (ra) lead.The physician suspected ra lead damage, however, diagnostic imaging was not performed.Abbott technical support was contacted and reprogramming was recommended.No intervention has been performed at this time.The patient was in stable condition and will continue to be monitored.