| Model Number |
CMRM6122 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Erosion (1750); Medical device site infection (4845)
|
| Date of Event |
08/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: this regulatory report is being submitted as part of a retrospective review and remediation related to capa 711948.Continuation of d10: 5076-45 lead implanted: (b)(6) 2015, 5076-52 implanted: (b)(6) 2015, w1dr01 ipg implanted: (b)(6) 2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced a pocket infection, erosion and pocket dehiscence.Its reported that the organism of infection was strep intermedius.The implantable pulse generator (ipg) system was removed and replaced.No further patient complications have been reported as a result of this event.It was further reported that an antibacterial absorbable envelope was placed during the implant procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected: d.3, d.4, h.2.4 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|