| Catalog Number |
L5C4531 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a homechoice device experienced a system error 2240 (air in line/set) alarm.The patient was connected at the time of the alarm.This occurred during drain two of four of peritoneal dialysis therapy.During troubleshooting, it was reported that there was a leak from the homechoice cassette that led to this alarm.Renal therapy services (rts) advised the patient to disconnect.When the patient took the cassette out of the cycler, the cassette had a couple drops of water.Rts advised the patient to drain manually.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h6 and h11: h11 : a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|