| Catalog Number |
MCS20 |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/6/2026 d4: batch # unk attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a neck lymph node dissection, clips were placed on a blood vessel, but they immediately dislodged.A second clip applier was opened, but the clips also dislodged.There was no blood loss.A reusable clip applier from another supplier was then opened, after which the clips remained in place and the procedure could be continued.The procedure was delayed by 5 minutes and no patient consequences were reported.The postoperative course was uncomplicated.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/7/2026.D4: batch # a9961p.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcs20 device was received with no apparent external damage.The opened packaging was also returned with the instrument.To replicate the reported incident, the device was subjected to functional testing.During testing, the device was cycled and successfully fed and formed the remaining fifteen (15) clips as intended.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Ensure that a clip is in the jaws.Position the jaws so that the clip completely surrounds the vessel to be ligated.Fully squeeze the handles together until they stop.As the handles are squeezed, the clip closes and occludes the vessel or tubular structure.Fully release the handles to return the instrument to its original open position.The next clip is automatically advanced.Inspect the jaws after each clip application to ensure the presence of a new clip before applying the next clip.Repeat steps as necessary to continue ligation of tissue.For optimal performance, periodically clean the instrument during the procedure by submerging the jaw and distal end of the applier shaft in saline.Device history review: a manufacturing record evaluation was performed for the finished device batch a9961p number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|