|
Block h6 device code a0401 is being used to capture the reportable event of suture break.Device evaluation block h11 the overstitch was not returned, and no media was available for analysis.The reported event of suture break could not be confirmed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on all gathered information, the probable cause selected is cause not established, because there is not enough evidence to determine whether the suture break was due to the physician's technique during the procedure or was related to a device malfunction.For the event additional device required, it happened during the procedure, and the most probable cause of this reported issue cannot be established due to lack of evidence.For this reason, will be classified as cause not established.
|
|
It was reported to boston scientific corporation that an overstitch 2-0 polypropylene suture was used during an esophagogastroduodenoscopy (egd) procedure on (b)(6) 2026.During the procedure, the suture broke.The physician attempted to use three additional sutures, which also broke.The procedure was completed with clips.There was no patient complications reported as a result of this event.Note: this is the third of four overstitch sutures used in the same procedure.
|