| Model Number |
DS760HS |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information regarding an bipap auto biflex device.There was no allegation of serious or permanent harm or injury.The device was returned to a third-party service center for evaluation.During evaluation, thermal damage to connector was found.No evidence of foam degradation was observed.
|
| |
|
Additional Manufacturer Narrative
|
|
On previously submitted report section additional narrative was incorrect.In this report updated/corrected as- the reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|