• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. AFFERA; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. AFFERA; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH Back to Search Results
Model Number AFR-00001
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a cardiac ablation procedure, when the catheter was removed from the right atrium to prepare for transseptal access, clots were noticed in the catheter lattice.Activated clotting time was measured at 337 at the time the catheter was removed from the body.The catheter was replaced.The case was completed.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the afr-00001 catheter with lot number 0232603181 was returned and analyzed.No anomalies were identified during external visual inspection of the catheter.During external visual inspection of the sphere segment, no anomalies were identified.The sphere was intact with no apparent issues.Tiny clot/blood like material was observed on the tip of the returned catheter.Data from the catheter identification chip confirmed that the device was used.The chip was re-programmed for functional testing.The returned catheter passed the mapping and ablation test with the mapping system (test/lab), no errors were triggered.No performance issues were identified with the returned catheter.Leak test was performed to check for any leaks on the irrigation line.The returned catheter passed the leak test.Occlusion test was performed to check for any blocks in the irrigation line.The returned catheter passed the occlusion test.Additional testing was performed where the catheter was connected to the irrigation pump and purged.Water could be seen dripping out of the nozzle, this was performed to confirm the irrigation line is not blocked or the nozzle holes.In conclusion, the reported issue (material in tip) was confirmed during device inspection.The catheter passed the returned product inspection.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFERA
Common Device Name
PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH
Manufacturer (Section D)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer (Section G)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key24799613
Report Number3012520654-2026-00149
Device Sequence Number13128913
Product Code QZI
UDI-Device Identifier00763000973391
UDI-Public(01)00763000973391(17)261206(10)0232603181
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P240013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAFR-00001
Device Catalogue NumberAFR-00001
Device Lot Number0232603181
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2026
Supplement Date Received by Manufacturer04/14/2026
Initial Report FDA Received Date04/07/2026
Supplement Report FDA Received Date05/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-