| Catalog Number |
IAB-05840-LWS-SC |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported " balloon did not inflate".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported " balloon did not inflate".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint for "balloon did not inflate" is not able to be confirmed.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action is required at this time.The reported complaint will be monitored for any developing trends.If the product is returned at a later date, a full investigation of the sample will be completed.
|
| |
|
Search Alerts/Recalls
|