ABBOTT MEDICAL TRICLIP¿ G5 STEERABLE GUIDE CATHETER; CATHETER, PERCUTANEOUS Back to Search Results
Model Number TSGC0801
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with tricuspid regurgitation (tr) for a triclip procedure.The procedure was completed successfully.After the procedure, it was difficult to remove the triclip steerable guide catheter (tsgc) from the stabilizer.Troubleshooting was performed, after significant force the sgc was removed from the stabilizer.The final tr was reduced.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
All available information was investigated, and the reported difficult to remove the tsgc from the stabilizer guide dock could not be replicated in a testing environment due to the returned condition of the device (damage pin hole of the guide attach).A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.The investigation determined that the reported difficult to remove may be related to a potential product quality issue.This complaint falls within the scope of an exception, as the complaint description and device code match the specific issue described in the exception.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICLIP¿ G5 STEERABLE GUIDE CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key24802120
Report Number2135147-2026-02214
Device Sequence Number10682161
Product Code DQY
UDI-Device Identifier05415067050724
UDI-Public(01)05415067050724(17)270607(10)51208R1004
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K243224
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTSGC0801
Device Catalogue NumberTSGC0801
Device Lot Number51208R1004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2026
Initial Report FDA Received Date04/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STABILIZER
Patient SexFemale
-
-