| Catalog Number |
STK-FE-001 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)4).
|
| |
|
Event or Problem Description
|
|
It was reported that an alert/notification settings issue occurred.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an alert/notification settings issue occurred.Product was provided for investigation.An external visual inspection was performed and failed due to usb connector damage.Receiver charge and boot tests were performed and failed.The speaker/vibrator resistance was measured and passed.An internal visual inspection was performed and passed.Performance data was reviewed.An alert/notification settings issue was not found within the investigation window.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|