• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL (CRM-SYLMAR) ALLURE RF; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL (CRM-SYLMAR) ALLURE RF; No Match Back to Search Results
Model Number PM3222
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/26/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found, as received, normal operating condition.Visual inspection of the header attachment area detected a bonding anomaly.The device was tested for a hermeticity breach which was not observed.The device was cut open to enable further testing and battery was operating above the elective replacement indicator (eri) voltage with appropriate remaining longevity.The hybrid circuitry was tested, and the results indicated normal current drain.Device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.
 
Event or Problem Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE RF
Common Device Name
No Match
Manufacturer (Section D)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24809228
Report Number2017865-2026-07197
Device Sequence Number10667347
Product Code NIK
UDI-Device Identifier05414734507905
UDI-Public(01)05414734507905(10)A000091738(17)210630
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2021
Device Model NumberPM3222
Device Lot NumberA000091738
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2026
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2026
Initial Report FDA Received Date04/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-