| Model Number |
PPLUS120 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint was reviewed and investigated according to the manufacturer¿s policy.Blocks a2-a6: not available from facility.Blocks b6 & b7: not available from facility.Block c: not applicable for this device.Block d4: serial # not applicable for this device.Blocks d6 & d7: not applicable for this device.Blocks h3 & h6: the pioneer plus catheter was discarded, thus no product investigation was performed.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
It was reported that a pioneer plus catheter was used for a peripheral procedure in severely calcified sfa.During use, the catheter was defective due to the ultrasound not working.Upon removal, the catheter became tangled/kinked, and the guidewire tore through part of the catheter; therefore, removed together.At this point, the physician decided it was better to discontinue the procedure and bring the patient back at a later date.No patient injury reported.This product problem is being submitted because the pioneer plus catheter and guidewire were removed as a system, and for delay of procedure.There is potential for harm if it were to recur.
|
| |
|
Search Alerts/Recalls
|