Model Number EB19-J10
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Video image failure(fluid damage).
 
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module with drive pcb.In addition, our technician confirmed that the insertion flexible tube crushed, the operation channel (primary) cut, and the operation channel (primary) resistance; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report (b)(4) (image failure) and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX MEDICAL
Common Device Name
VIDEO BRONCHOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
paresh patel
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key24812456
Report Number9610877-2026-50765
Device Sequence Number9594188
Product Code EOQ
UDI-Device Identifier04961333221530
UDI-Public(01)04961333221530(11)211021
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K200678
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEB19-J10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2026
Initial Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-