| Model Number |
SIGTRSB60AMT |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
03/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, in a laparoscopic gastric bypass sleeve, during the transection of the stomach, the powered stapler partially fired.The firing stop half way through the firing.The staple line was resected and re-stapled due to the issue.The issue happened with four reloads.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sigadaptxl, sig power sigadaptxl linear xl adapter, (serial # (b)(6)); sigtrsb60amt, sigtrsb60amt 60mm med thk reinforced rel, (lot # n5l1164y); sigtrsb60amt, sigtrsb60amt 60mm med thk reinforced rel, (lot # n5l1164y); sigtrsb60amt, sigtrsb60amt 60mm med thk reinforced rel, (lot # n5l1164y); sigphandle, sig power sigphandle handle, (serial # unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.Six representative samples were available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device experienced partial firing.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|